Place of Origin: | China |
Brand Name: | HUGE OEM ODM |
Certification: | CE RoHS |
Model Number: | CONTEC8000A |
Minimum Order Quantity: | 1PC |
---|---|
Price: | US $ 340-400 / PC |
Packaging Details: | box |
Delivery Time: | 5-10days |
Payment Terms: | TT paypay wester union |
Supply Ability: | 100000 PCS per Month |
Model:CONTEC8000A
Introduction
CONTEC8000 ECG workstation has such functions as 12-lead and 3-lead synchronous ECG collection, replay, automatic analysis and diagnosis, spectrum ECG, High Frequency ECG, QT internal Dispersion analysis, Heart Rate Variability (HRV) analysis, myocardial ischemia analysis, heart rate vibration analysis, pacing ECG analysis, Vector Cardiogram (VCG), Time Vector Cardiogram (TVCG), SAECG analysis etc.These functions can reflect the heart's locomotion circumstance and provide credible gists for doctors to diagnose.
Main Features
*Clear collecting template, instant sampling, display and saving. Instant QRS analysis, HR computation and real time printing. DFT/AC/EMG filters.
*Advanced ECG parsing arithmetic is adopted in the software, so that it is more exact when the computer is seeking points, and it can obtain reliable auto-recognizing sort of arrhythmia, which can make the analyzing report and conclusion more valuable.
*Lead position adjustment, solves the problem of wave over-lapped, caused by baseline shift or too high wave crest, and can arrange the lead position automatically.
*Bran-new orientating diagnosis for coronary heart disease is finer to analyze abnormal Q wave.
*The QTD template is provided to display the Chromatogram Map. It can also show the 12-lead ECG waves in high fidelity in order to ensure the adjustment of the points’ position more exactly.
*HRV supports the analysis and printing of the long case. You can press left button of the mouse to select the analyzing time conveniently.
*It can analyse heart rate vibration, myocardial ischemia, pacing ECG simultaneously.
*The shortcut replay function of tridimensional vector loop can make you review the trend of vector loop clearly.
*The Print Preview and Print functions of "What You Get Is What You See" can be found in every template. The A4, B5 two sizes of paper are supported, so you can use it conveniently.
*The "shortcut print" function can help you preview or print all of the analytical reports quickly, and save the reports in your computer with multimedia format(PDF format JPG format BMP format or PNG format) according to factual need , so that you can turn over them more conveniently and intuitively in future.
*High-sampling rate (1000Hz) ensures ECG waves not anamorphic, and fine resolution in display and print , shows details of ECG waves perfectly.
*Strong and perfect base management of case makes it convenient to exceed, lookup, delete, backup, recover and modify the case history data, and supports the combined conditions query function.
*Abundant setting items are provided in system setting, where you can set basic information, the operation of software, data sampling, wave analysis, database management, printing etc, according to your special need.
Performance
HR display range is 30bpm300bpm, and error is the larger of ±1% and ±1bpm.
Sampling frequency:1000Hz.
Sampling precision:12bits.
Lead:standard 12 leads and 3 leads.
Input mode: floating and defibrillating protection.
Input impedance:≥5MΩ.
Minimum test signal:the sine signal of 10Hz, 20μV.
Input Loop current:≤0.1μA.
Dynamic input range:±5mV.
Time constant:≥3.2 s.
CMRR:≥80dB.
50Hz interference suppression:≥20dB.
Amplitude frequency characteristic:0.5Hz150Hz(+0.4dB -3dB).
Gain:51020mm/mV,error: ±5%.
Bear Polarizing voltage: ±300mV, Sensitive degree change ±5%.
Noise Voltage(P-P):≤30μV.
ECG sampling box(1)
ECG lead wire(1)
Grounding Cable(1)
Limb electrode holder(4)
Chest electrode absorbing ball(6)
CONTEC8000 ECG workstation installation CD(1)
ECG lead extension cable(transfer electrode film)(2)
System configuration of host
CPU: above 1.2GHz
Memory: above 128M
Mainboard: supporting USB2.0 device, the mainboard with Intel chip group is recommended.
HD: above 10G
CD ROM: CD-ROM with 24 times speed or above (or CD memory System)
Operating System:Microsoft Windows 2000/XP/Vista(XP is recommended)
Display resolution: above 1024*768
Interface: USB2.0